The sensitivity calibration was performed in air for VIS, in air and vacuum for MUV, and in vacuum for FUV. The VIS and MUV experiments performed in air were done with the instrument pointed at a reflectance screen with the lamp at varying distances and for different lamps. After data was taken, the direct illumination of the screen was occulted and then data was taken again to account for stray light in the lab. Next the telescope was closed and data taken in order to account for dark counts with the instrumemt. These last two data sets were subtracted off from the original data set. For the MUV and FUV vacuum experiments, collimated light from a deuterium lamp point source was sampled by both the instrument and calibrated detectors. Note the difference between "MUV_sensitivity_Moby2" and "MUV_sensitivity_Moby3" is that the instrument was rotated 90 degrees about the optical axis, and similarly for "FUV_sensitivity_Moby2" and "FUV_sensitivity_Moby3" file environment lamp lamp-screen distance instrument input channel v_sen_fel_15a air fel 197.5 cm illuminated screen VIS/MUV v_sen_fel_15b air fel 197.5 cm illuminated screen VIS v_sen_fel_dark_15a air fel 197.5 cm telescope closed VIS/MUV v_sen_fel_dark_15b air fel 197.5 cm telescope closed VIS v_sen_fel_shutter_15a air fel 197.5 cm screen occulted VIS/MUV v_sen_fel_shutter_15b air fel 197.5 cm screen occulted VIS V_sen_fel_15c air fel 392.75 cm illuminated screen VIS v_sen_fel_15d air fel 392.75 cm illuminated screen VIS v_sen_fel_dark_15c air fel 392.75 cm telescope closed VIS v_sen_fel_dark_15d air fel 392.75 cm telescope closed VIS v_sen_fel_shutter_15c air fel 392.75 cm screen occulted VIS v_sen_fel_shutter_15d air fel 392.75 cm screen occulted VIS v_sen_deu_15a air deuterium 99.0 cm illuminated screen VIS/MUV v_sen_deu_shutter_15a air deuterium 99.0 cm screen occulted VIS/MUV MUV_Sensitivity_Moby2 vacuum deuterium ------ collimated source MUV MUV_Sensitivity_Moby3 vacuum deuterium ------ collimated source MUV FUV_sensitivity_Moby2 vacuum deuterium ------ collimated source FUV FUV_sensitivity_Moby3 vacuum deuterium ------ collimated source FUV