For a detailed description of the TLS instrument in SAM, the reader is referred to P. R. Mahaffy et al., "The Sample Analysis at Mars Investigation and Instrument Suite", Space Sci. Rev. 170, 401-478 (2012). For detailed description of the spectral scans, line identification, data processing and calibration, the reader is referred to the manuscripts and Supplementary Online Material (SOM) accompanying two Science papers: (i) C.R. Webster et al., "Isotope Ratios of H, C and O in CO2 and H2O of the Martian Atmosphere", Science 341, 260 (2013); and (ii) C.R. Webster et al., "Low Upper Limit to Methane Abundance on Mars", accepted for publication in Science, 2013. These latter two papers focus on atmospheric data and results, while for the Evolved Gas Analysis (EGA) runs, further details are given in a companion Science paper authored by Laurie Leshin et al. in the same volume. Briefly, for EGA analysis, ingested solid sample is heated in a pyrolysis oven whose gaseous products are routed to either QMS, TLS or GC in a helium gas flow. For any EGA run, TLS is given a "temperature cut" where the effluent is delivered to the TLS Herriott cell during a certain oven temperature portion of its heating ramp, typically a 50-100 deg C cut that may be at lower or higher temperature, and identified in our level 2 data set. Isotopic ratios for d17O and d13C18O, while provided for standard "atmospheric" runs, cannot retrieve meaningful values for EGA runs where spectral signals (line depths) are much weaker. Rocknest EGA effluent gases were observed by the QMS to contain contamination from a derivatization agent MTBSTFA that was taken along with the SAM experiment. For TLS, numerous underlying spectral lines were observed in our spectral scans whose contribution to the CO2 and H2O isotopic lines could not be unambiguously identified. However, these interferences were generally small, and diminished throughout the Rocknest run series and made only minor contributions to the John Klein and Cumberland analyses. For EGA analysis, it is important to take into account what gases and isotope ratios would result with NO sample in the pyrolysis oven. To that end, the EGA run is duplicated in this condition (no sample) and minor amounts of both CO2 and H2O are produced and detected in TLS. Because these "blank cup" gases are measured by TLS to have isotopic ratios different from the Mars solid samples, we have made corrections to the level 2 data based on a blank cup analysis usually run immediately before each series (Rocknest, John Klein, Cumberland). These corrections are described in Leshin et al. Science, submitted 2013. Isotope ratios given refer to SMOW for oxygen isotopes and VPDB for d13C ratios. These ratios were determined using identical data processing software in comparison to certified calibration gas or liquid standards determined by IRMS (see published SOM). All isotope delta values and their standard errors (one-sigma, SE) are given in "per mil" according to standard convention. In some cases (entries marked with "x") scientifically useful data was not retrieved due to instrumentation issues such as fast moving spectral scans or poor signal-to-noise ratios. In atmospheric cases, values for d13C and d18O result from the mean value from the two spectral regions 1 and 2 accessed by a single laser by changing the laser temperature, but in the EGA results, we use line-pairs that appear better behaved with respect to contaminant or underlying spectral line issues. We emphasize that the understanding and analysis of identified and unidentified contaminants and underlying spectral lines is ongoing, and will result in future corrections (updates) to our level 2 data. *** This is the second Cumberland sample analysis.