The retention times are calculated from the moment the sample is injected in the chromatographic column (IT FLASH or GC INJECTION). In the chromatogram, this corresponds to the first peak observed in time, which is generated by the flash heating of the injection trap of SAM-GC. This time is 534788735.12 s (sclk) for GC 1 chromatogram and 534793066.5 s (sclk) for GC 4 chromatogram . The retention times are calculated by subtracting this time, to the times of the peaks maxima. The original TCD signal is available in the file noise.csv. GC column 1 run peak 0 ; GC1 INJECTION ; 0 peak 1 ; ; 404 peak 2 ; ; 498 peak 3 ; ; 614 peak 4 ; ; 842 peak 5 ; ; 1049 peak 6 ; ; 1102 peak 7 ; ; 1969 GC column 4 run peak 8 ; GC4 IT FLASH 1 ; 0 peak 9 ; ; 178 peak 10 ; ; 224 peak 11 ; ; 263 peak 12 ; ; 685 peak 13 ; BENZENE ; 701 peak 14 ; TOLUENE; 900 peak 15 ; ; 1019 peak 16 ; ; 1061 peak 17 ; ; 1087 peak 18 ; GC4 IT FLASH 2 ; 1101 peak 19 ; ; 1109 peak 20 ; ; 1287 peak 21 ; ; 1319 peak 22 ; ; 1325 peak 23 ; ; 1326 peak 24 ; ; 1345 peak 25 ; ; 1380 The abundance given in the table (species.csv) corresponds to the area of the most abundant mass peak which are measurable for each peak. The unit is counts/sec/sec. There is no baseline correction for this fit. These surfaces are proportional to the amount of the species detected when a peak is eluted. When peaks are co-eluted (over of different peaks), the peaks are fitted using a gaussian function. If the fit does not work, only one surface value is given, corresponding to the sum of the abundances of the species contributing to the peak.