The retention times are calculated from the moment the sample is injected in the chromatographic column (IT FLASH or GC INJECTION). In the chromatogram, this corresponds to the first peak observed in time, which is generated by the flash heating of the injection trap (IT) of SAM-GC. This time is 656587830.42 s (sclk) for GC 1 chromatogram and 656591804.32 s (sclk) for GC 2 chromatogram. The retention times are calculated by subtracting this time, to the times of the peaks maxima. The original TCD chromatograms are available in the file noise.csv. GC column 1 run peak 0 ; GC1 INJECTION ; 0 peak 1 ; ; 319 peak 2 ; ; 541 peak 3 ; ; 713 peak 4 ; ; 766 peak 5 ; ; 900 peak 6 ; ; 1722 GC column 2 run peak 7 ; ; -653 peak 8 ; GC2 INJECTION ; 0 peak 9 ; ; 814 peak 10 ; ; 1176 peak 11 ; ; 1588 The abundance given in the table (species.csv) corresponds to the area of the most abundant mass peak which are measurable for each peak. The unit is counts/sec/sec. There is no baseline correction for this fit. These surfaces are proportional to the amount of the species detected when a peak is eluted. When peaks are co-eluted (over of different peaks), the peaks are fitted using a gaussian function. If the fit does not work, only one surface value is given, corresponding to the sum of the abundances of the species contributing to the peak.