This experiment was a blank run of the drilled solid sample experiment sequence. It revealed the continued presence of products of the derivatization reagent, N-methyl-N-tert-butyldimethylsilyl-trifluoroacetamide (MTBSTFA), adsorbed onto the sample. For estimation of abundances of volatile compounds detected in EGA, the background was subtracted as a constant value, either at the start or end of the run or before major peaks. An exception to this was CO2, which was modeled with a downward-trending background throughout the run. Abundances are reported only for gases evolving during the pyrolysis ramp and not the preheat, in keeping with data for EGA runs. Thus there are no abundances reported for H2O or HCl, even though they were present in the preheat. Uncertainties in abundances, deriving from statistical noise, background subtraction,calibration factors, assumptions regarding ionization cross-sections for the relevant compounds, and isobaric interferences, are generally estimated at 10-20% of the values.