Reconciled with block: tega_ega_whack

Objectives: Switch from Filament 0 to Filament 1, with a short on Filament 0. Run sweep to evaluate quality of data in this state.

Configuration: TEGA performed Checkout I – successfully. Checkout II results indicated ion source short on Filament 0.

Filaments used: 1

SW Version (386): SW Version (EGA):

Sweep and hop mode version(s): V112

Super modes and Sweep modes used: Sweep 38

Block duration: 25 minutes

1) TEGA ON
2) Select oven (as input parameter)
3) Sequence 30A
4) EGA On
a) Restore modes EEprom 0
5) Set EGA Transmit time to 1 sec
6) Set Jplates (-107,-92V)
7) Select Filament 1
8) Start SM 1
9) MV ON @ 3200V
10) Ion Pump ON
11) Emission ON
12) Pause 2 minutes
13) Stop SM 1
14) Pause 3 minutes
15) Run Sweep 39 (long version)
16) Turn EGA Off
17) Turn TEGA Off