Objectives:
This procedure checks that the instrument is still in good operational condition. It will also get a background measurement at the end. Block duration written not to exceed 4 hours.
Configuration: Landed Operations: Surface Phase Characterization. Checkout I & Checkout II_Puncture should have been executed and instrument nominal.
Assumptions: TEGA can not be turned on below -50C
Gas(es) used: Calibration and Carrier Gas Tanks have been punctured. No gases used during this activity.
Filaments used: 0
SW Version (386): V513 SW Version (EGA): V321
Sweep and hop mode version(s): V111
Super modes and Sweep modes used: SM 3, Sweep Modes 37
Sequences used: 30A,26A
Sets Jplate 1 to -92V
Start SM 1 pause 1 sec
Set MV 2800V Pause 1 sec
Ion Pump ON Pause 1 sec
Emission ON
Pause 2 Minutes
Stop SM 1