Reconciled with block: tega_door_open\r26
Objectives: Open the TA Door and collect data that permits a calculation of the heat capacity of the female oven half. This version is to be used for TA#4 – Surface Phase Characterization ONLY.
Configuration: This procedure may begin with TEGA off, or it may be tacked on the end of some other procedure.
Filaments used: 0
SW Version (386): SW Version (EGA):
Sweep and hop mode version(s): None
Super modes and Sweep modes used: None
-
1) TEGA ON (called in interop block)
-
2) Select oven (as input parameter) Sol _____ TA _______
-
3) Command TA Door Open
-
a) Open door default duration is 5 minutes
-
b) GV_TEGA_TA_DONE timeout is 10 minutes. If timeout does not go true the activity is terminated.
Heat manifolds to ensure pressure sensors work.
-
4) Start TA Manifold Heater to regulate at 20 C [tweakable]
-
5) Start EGA Manifold Heater to regulate at 35 C
-
6) Pause 5 minutes for manifolds to come to temperature
Exercise valves and carrier gas to check out flow and ensure no clogged frits.
-
7) Call tega_flow_car (executes Seq 90A open/close of carrier valve)
-
8) Pause 1 minute (to measure leaks)
-
9) Open bypass valve
-
10) Pause 1 minute (to check out EGA pressure sensor)
-
11) Close bypass valve
-
12) Open oven inlet valve
-
13) Pause 1 minute (to measure flow rate to ensure frit is not plugged)
-
14) Close oven inlet valve
-
15) Set Carrier gas to off: Calls tega_stop_car
-
16) Open oven outlet valve
-
17) Pause 1 minute (to empty gas from TA manifold)
-
18) Close oven outlet valve
-
19) Stop TA manifold heater regulation
-
20) Turn manifold heater off
-
21) Stop EGA manifold heater regulation
-
22) Turn EGA manifold heater OFF
Check out oven fill and close oven for copper path
-
23) Start Oven fill – Shake for 15 secs (verifies LED operational)
-
a) Will not shake/LED Diff Parameters set to prevent
-
24) Close Oven – 10 secs
-
a) This has a 30 sec timeout for GV_TEGA_TA_DONE. IF GV does not go true, activity gracefully terminated.
-
25) Ramp oven as follows: Call tega_ramp (hidden child)
-
a) Ramp Segments: TA#4 = 0-3A (PIT values are different than Flight Values)
-
i) -50 to -30 at 20 K/min
-
ii) Hold at -30 for 1 minute
-
iii) -30 to -10 at 5 K/min
-
iv) Hold at -10 for 1 minute
-
v) -10 to -30 at -5 K/min
-
26) Shut down (call tega_power_off)