November 10, 2008 Revision 2
TEGA Surface Phase Characterization Sample Acquire
Objectives:
This procedure turns on TEGA, prepares the instrument for the initial surface phase characterization sample, acquires sample delivered from the RA and closes the oven to TA #4.
Configuration: Landed Operations: Surface Phase Characterization. Instrument has gone through Checkout I, Checkout II Puncture seals, Cover retract has been complete and TA #4 door open has been completed and verified.
Assumptions: TEGA can not be turned on below -50C
Gas(es) used: Calibration and Carrier Gas Tanks have been punctured. No gases are flown during this activity. Filaments used: 0 SW Version (386): V513 SW Version (EGA): V321 Sweep and hop mode version(s): V112 Super modes and Sweep modes used: None. The EGA is not ON during this activity.
3.1.1. Call tega_select_oven (TWEAKABLE – NEED TO VERIFY TA#)
- 1 - November 10, 2008 Revision 2
TEGA Surface Phase Characterization Sample Acquire
| 3.1.2. | Start solenoid shaker |
| 3.1.3. | Wait for oven Full: GV_TEGA_OVEN_FULL has a timeout of 15 minutes |
| 3.1.4. | Close Oven – wait for GV_TEGA_TA_DONE |
| 3.1.5. | Oven close time out is 20 minutes |
4. Call tega_power_off Dump FPGA Registers
4.1. Main 3 Off
4.2. Main 2 Off
4.3. Main 1 Off
- 2 -