June 5, 2008 TEGA_Mid-Temp_v4_r3.doc
Objectives:
Power on and perform the medium-temperature ramp on a soil/ice sample that had been previously heated via a low-temperature ramp. The purpose of this run is mainly to help decompose any possible oxidizing agent that might be on the soils. It also looks for gas release on the way up to the hold temperature. After warm up, we will collect background data for 30 minutes then begin the ramp. The ramp starts at -60 C and runs to + 175 C at 20 K/min followed by a 60 minute hold to measure the D/H ratio on any H2O that may be evolved.
This procedure also checks for leaks in the manifold, leaks in the oven, and checks the pressure in the supply tanks (simply by looking at the step rise in pressure with each pulse).
Note this run will be entirely at low emission.
Note also that this run has no mass-to-voltage calibration after the one done right after warm up.
Configuration:Low-temperature ramp has been completed previously. Constraints:
Review data from Checkout II to confirm dN numbers for regulation for in activity block and ramp segments.
Review: LT Results. Did the water evaporate enough or do we need to repeat LT? _ __
Gas(es) used: Carrier and Calibration. Mars atmosphere will also be collected through the TA Sample valve. Filaments used: 1 LOW Emission SW Version (386): V513 SW Version (EGA): V321 Sweep and hop mode version(s): MODE TABLE V117.
Super modes and Sweep modes used: Supermodes 1, 2, 3, 6, 7. and 10; Sweeps 37, and 38. Supermodes running in ramp: 6, 7 & 10
Note: TEGA is checked to see if TEGA is safed. If so, abort self to end activity
- 1 -
June 5, 2008 TEGA_Mid-Temp_v4_r3.doc
TEGA Mid-Temp Soil
6.1.1. Note: IF pass through was selected as HIGH, calibration set CANNOT equal 3
6.2. Calls tega_ega_prep, filament-16.2.1.1.Sets JPlate 0 to -107V 6.2.1.2.Sets Jplate 1 to -92V6.2.1.3.Start SM 1 pause 1 sec 6.2.1.4.Set MV 2800V Pause 1 sec6.2.1.5.Ion Pump ON Pause 1 sec6.2.1.6.Emission ON6.2.1.7.Pause 2 Minutes6.2.1.8.Stop SM 1
6.3. Change engineering transmit to 10 secs
6.3.1. IF pass thru = 1, emission current will change to HIGH
6.4. Call tega_ega_calibration_scans (emission low, cal set 3, 3200V)
6.4.1. MV exits at 3200V
6.5. Wait 1 Minute
6.6. Start SM 3
6.7. Pause 20 minutes
6.8. Stop SM 3
- 2 - June 5, 2008 TEGA_Mid-Temp_v4_r3.doc
TEGA Mid-Temp Soil
6.9. Pause 1 minute
6.10. Calls tega_ega_gain
6.10.1.1.1. Start SM 2
6.10.1.1.2. Pause 1 minute
6.10.1.1.3. Set MV 2600 – 3400V (1 minute per voltage)
6.10.1.1.4. Remain at 3200V
6.10.1.1.5. Stop SM 2
6.10.1.1.6. Run Sweep 38
EGA is warmed up and ready to go (54 mins)
Execute sequence 24A
Execute Sequence 24 B to increase readings on temperature regulation.
14.1.1. TA Manifold +65C
14.1.2. Plumbing 1 +35C
14.1.3. Plumbing 2 +35C
14.1.4. Transfer Tube +75C
14.1.5. EGA Plumbing +45C
14.1.6. EGA manifold to +35 C
14.1.7. Wedge Heater @ 10% duty cycle
- 3 -
June 5, 2008 TEGA_Mid-Temp_v4_r3.doc
TEGA Mid-Temp Soil
27.1.1. SM 6 running
27.1.2. Carrier and Calibration Closed
27.1.3. Inlet & Outlet closed
27.1.4. TA Sample Closed
27.1.5. Bypass OPEN
Shut down
32.1.1.1. EGA Shutdown:
32.1.2. Multiplier Voltage OFF
32.1.3. Wait 5 secs
32.1.4. Emission Off
32.1.5. Wait 5 secs
- 4 -
June 5, 2008 TEGA_Mid-Temp_v4_r3.doc
TEGA Mid-Temp Soil
32.1.6. Ion Pump Control OFF
32.1.7. Activate Sequence 7A to turn the EGA OFF
33. Turn off TEGA: CALL tega_power_off
33.1.1. Turn off the third TEGA switch
33.1.2. Turn off the second TEGA switch
33.1.3. Turn off the first TEGA switch
33.1.4. FPGA Dump – Aker Valve state
- 5 -