Objectives: The purpose of this block is to run Cal Gas and Carrier Gas. The Cal Gas will be run at both high and low emission to calibrate past (and future) TEGA runs. The run will emphasize getting data on CO2 isotopes and D/H ratio. We will also spend a few minutes collecting some data to help understand some peculiar things noticed in hops not being reproducible and we will spend some time conditioning channel 2 with O+ ions.
Configuration: Instrument is in its normal configuration.
Gas(es) used: Carrier and Calibration.
Filaments used: 1, High and Low Emission
Sweep and hop mode version(s): MODE TABLE V122.
Super modes and Sweep modes used: Supermodes 1, 2, 3, 4, 5, 11, 12, 13, & 15; sweeps 37, 38.
Note: TEGA is checked to see if TEGA is safed. If so, abort self to end activity
6.1.1. Note: IF pass through was selected as HIGH, calibration set cannot equal 3
- 1 -
July 5, 2008 TEGA_Cal-Gas_only_v1_r1.doc
TEGA Cal-Gas Run
6.2. Calls tega_ega_prep, filament-16.2.1.1.Sets JPlate 0 to -110V 6.2.1.2.Sets Jplate 1 to -92V6.2.1.3.Start SM 1 pause 1 sec 6.2.1.4.Set MV 2800V Pause 1 sec6.2.1.5.Ion Pump ON6.2.1.6.Pause 30 seconds6.2.1.7.Emission ON6.2.1.8.Pause 2 Minutes6.2.1.9.Stop SM 1
6.3. Change engineering transmit to 10 secs
6.3.1. IF pass thru = 1, emission current will change to HIGH
6.4. Call ega_calibration_scans (Set =3, exit = 3200V)
6.5. Wait 2 Minutes
6.6. Start SM 3
6.7. Pause 20 minutes
6.8. Stop SM 3
6.9. Pause 1 minute
6.10. Calls tega_ega_gain
6.10.1.1.1. Start SM 2
6.10.1.1.2. Pause 1 minute
6.10.1.1.3. Set MV 2600 – 3400V (1 minute per voltage)
6.10.1.1.4. Remain at 3200V
6.10.1.1.5. Stop SM 2
6.10.1.1.6. Run Sweep 38
EGA is warmed up and ready to go (54 mins)
Measure residuals and get set up
- 2 -
Note different temperatures below
Execute Sequence 24 B to increase readings on temperature regulation.
14.1.1. Plumbing 1 +35C
14.1.2. Plumbing 2 +35C
14.1.3. Transfer Tube +35C
14.1.4. EGA Plumbing -10 C
14.1.5. EGA manifold to -10 C
14.1.6. Wedge Heater @ 10% duty cycle
15. Pause 5 minutes for residuals at low emission
16. Switch to high emission
Start flowing Cal Gas
- 3 - - 4 -
48. Run sweep mode 41 (Ar Peak)
Flow carrier gas to blow Cal Gas out of CML volume mainly to determine arrival time from manifold to EGA.
Shut down
65.1.1.1. EGA Shutdown:
65.1.2. Multiplier Voltage OFF
65.1.3. Wait 5 secs
65.1.4. Emission Off
- 5 -
65.1.5. Wait 5 secs
65.1.6. Ion Pump Control OFF
65.1.7. Activate Sequence 7A to turn the EGA OFF
66. Turn off TEGA: CALL tega_power_off
66.1.1. Turn off the third TEGA switch
66.1.2. Turn off the second TEGA switch
66.1.3. Turn off the first TEGA switch
66.1.4. FPGA Dump – Aker Valve state
- 6 -