Aug 17, 2008 TEGA_abnormal_Atmosphere_open_carrier_r2(s109).doc

TEGA Low-Emission Atmosphere Run Reconciled with block: tega_atmos_r33_s109 Objectives: The purpose of this block is to open the carrier valve, test regulating carrier using the inlet valve, and measure the composition of the carrier gas

Configuration: Instrument is in its normal configuration. Constraints: Mode table V123 must have been successfully loaded Gas(es) used: carrier Filaments used: 1, Low Emission SW Version (386): V513 SW Version (EGA): V321 Sweep and hop mode version(s): MODE TABLE V123. Super modes and Sweep modes used: Supermodes 1, 2, 3, 6; sweeps 37, 38 Inputs: Filament: 1 High_current: 0 Calibration_set: 3

Channels: 0x9

Mult_volt: 3300

Four_volt: 2800

Pump_nudge: 0

Note: TEGA is checked to see if TEGA is safed. If so, abort self to end activity

1. Turn ON TEGA – Call TEGA_power_on

1.1.1. TEGA_MAIN1_PWR "ON"
1.1.2. TEGA_MAIN2_PWR "ON"
1.1.3. TEGA_MAIN3_PWR "ON"
1.1.4. Start the TEGA 386 FSW

2. Preheat for as long as possible (30 seconds minimum

3. Turn on EGA

3.1.1. TEGA/FSW:EGA:Change EGA State
3.1.2. Restore the modes from EEPROM 0
3.1.3. EGA Commands:Restore Mode
3.1.4. Run mode 37

- 1 -

Aug 17, 2008 TEGA_abnormal_Atmosphere_open_carrier_r2(s109).doc

TEGA Low-Emission Atmosphere Run

  1. OPEN GEC Vent Valve
  2. Warm up EGA: CALL tega_ega_emission_warmup (filament 1, calibration_set, high_current, mult_volt, four_volt, channels) 6 inputs
5.1.1. Note: IF pass through was selected as HIGH, calibration set cannot equal 3

5.2. Calls tega_ega_prep, filament-15.2.1.1.Sets JPlate 0 to -107V 5.2.1.2.Sets Jplate 1 to -92V5.2.1.3.Start SM 1 pause 1 sec 5.2.1.4.Set MV 3300V Pause 1 sec5.2.1.5.Ion Pump ON5.2.1.6.ause 1 second 5.2.1.7.Emission ON5.2.1.8.Pause 2 Minutes5.2.1.9.Stop SM 1

5.3. If high_current = 1, emission current will change to HIGH

5.4. Change engineering transmit time to 10 seconds

5.5. Call ega_calibration_scans (Set =3, exit = 3300V) inputs calibration_set, high_emission, channels, scan_voltage, chan_four_voltage, exit_voltage

5.6. Start SM 3

5.7. Pause 20 minutes

5.8. Stop SM 3

5.9. Pause 1 minute

5.10. Calls tega_ega_gain

5.10.1.1.1. Start SM 2
5.10.1.1.2. Pause 1 minute
5.10.1.1.3. Set MV 2600 – 3400V (1 minute per voltage)
5.10.1.1.4. Remain at 3300V
5.10.1.1.5. Stop SM 2
5.10.1.1.6. Run Sweep 38

EGA is warmed up and ready to go (54 mins)

(no pump nudge)

Measure residuals and get set up

- 2 -

Aug 17, 2008 TEGA_abnormal_Atmosphere_open_carrier_r2(s109).doc

TEGA Low-Emission Atmosphere Run

  1. Change TA_manifold and Outlet Pressure to timer1
  2. Change bus A and bus B current to timer 1
  3. Execute Sequence 24 B to increase readings on temperature regulation.
  4. Regulate TA manifold to +65C
  5. Regulate EGA manifold to +35C
  6. Calibrate EGA: CALL tega_ega_calibration_scans (set = 3, exit=3300 V) ) inputs calibration_set, high_emission, channels, scan_voltage, chan_four_voltage, exit_voltage
  7. Start Supermode 6
  8. Select oven for gas flow – TA0 (defaults to oven 0)

Briefly measure atmosphere on TA inlet

  1. Open TA Sample valve (to measure atmosphere)
  2. 15. Pause 1 5 minutes

Regulate whatever is in the manifold through TA 0

    1. Loop 15 times (should be 5 minutes total)
    2. 16.1. open inlet valve
    3. 16.2. pause 5 seconds
    4. 16.3. close inlet valve
    5. 16.4. pause 15 seconds
  1. 17. Close TA sample valve
  2. Open Outlet valve

- 3 -

Open the carrier valve

  1. Set valve pulse time to 4 mS
  2. 20. Loop 200 times (should take 17 minutes)

Aug 17, 2008 TEGA_abnormal_Atmosphere_open_carrier_r2(s109).doc

TEGA Low-Emission Atmosphere Run

20.1. Open carrier valve

20.2. pause 5

Regulate carrier gas with the inlet valve and measure gas composition

  1. Open TA sample valve
    1. loop 15 times (should be 5 minutes)
    2. 22.1. open inlet valve
    3. 22.2. pause 5 seconds
    4. 22.3. close inlet valve
    5. 22.4. pause 15 seconds
  2. pause 5 minutes to watch pressure rise

Clean up and shut down

  1. Close TA Sample valve
  2. Stop SM6
  3. Disable Heaters
  4. Close Outlet valve
  5. 28. CLOSE GEC Vent Valve
    1. Turn off EGA: CALL tega_ega_off
    2. 29.1.1. EGA Shutdown:
    3. 29.1.2. Multiplier Voltage OFF
    4. 29.1.3. Wait 5 secs
    5. 29.1.4. Emission Off
    6. 29.1.5. Wait 5 secs
    7. 29.1.6. Ion Pump Control OFF
    8. 29.1.7. Activate Sequence 7A to turn the EGA OFF
  6. Turn off TEGA: CALL tega_power_off

30.1.1. Turn off the third TEGA switch

30.1.2. Turn off the second TEGA switch

30.1.3. Turn off the first TEGA switch

30.1.4. FPGA Dump – Aker Valve state