Reconciled with block: tega_door_open\r36_TA1_6
Objectives: The Carrier valve since a sol 111 has been consistently opening on command. This block provides a confidence test for valve opening and closing on command. The final state of the carrier valve shall be in the “Open” state. It will also do a re-check of our ability to regulate gas via the bypass and oven inlet valves. (The test on Sol 118 gave a too high pressure)
Configuration: Needs to be run other than first thing in the morning because we don’t want this run at a lower PEB temp of -40. This procedure may begin with TEGA off, or it may be tacked on the end of some other procedure.
TA Manifold Pressure is collected on Timer 1
Filaments used: None
SW Version (386): 515 SW Version (EGA): N/A
Sweep and hop mode version(s): None
Super modes and Sweep modes used: None
-
1) TEGA ON
-
2) Execute Sequence 30A
-
3) Select oven (as input parameter) Sol _121____ TA ___0____
-
4) Activate Sequence 24A
-
5) Change valve Pulse time to 4mS
-
6) Set TA_MANIFOLD_PRES to timer_1
-
7) Set TA OUTLET_PRES to timer 1
-
8) Loop 100 times (10 second loop)
-
a) Drive valve open using register write =2049
-
b) FPGA Dump B to verify valve state.
-
c) Drive valve closed using register write =4097
-
d) FPGA Dump B to verify valve state.
-
9) Start TA Manifold Heater to regulate at 65 C (Heat manifolds to ensure pressure sensors work
-
10) Start EGA Manifold Heater to regulate at 35 C
-
11) Loop 100 times (10 second loop)
-
a) Drive valve open using register write =2049
-
b) FPGA Dump B to verify valve state.
-
c) Drive valve closed using register write =4097
-
12) FPGA Dump B to verify valve state Loop 180 times (5 second loop) aka “A BUNCH”
-
a) Drive valve open using register write =2049
-
b) FPGA Dump B to verify valve state.
-
13) Set timer 5 to 8 s (to control on-off times of inlet or bypass valves for flow regulation)
-
14) Control bypass valve via duty cycle
-
15) Pause 520 seconds
-
16) Stop control of bypass valve
-
1. Set timer 5 to 16 s (to control on-off times of inlet or bypass valves for flow regulation)
-
17) Open oven outlet valve
-
18) Control oven inlet valve via duty cycle
-
19) Pause1040 seconds
-
20) Stop control of inlet valve
-
21) Close outlet valve
-
22) Delay 10 minutes to determine leak rate
-
23) Disable heaters
-
24) Do TEGA TA 4 check with “check_oven=0”
-
25) Call power off