Reconciled with block: tega_door_open\r36_TA1_6

Objectives: The Carrier valve since a sol 111 has been consistently opening on command. This block provides a confidence test for valve opening and closing on command. The final state of the carrier valve shall be in the “Open” state. It will also do a re-check of our ability to regulate gas via the bypass and oven inlet valves. (The test on Sol 118 gave a too high pressure)

Configuration: Needs to be run other than first thing in the morning because we don’t want this run at a lower PEB temp of -40. This procedure may begin with TEGA off, or it may be tacked on the end of some other procedure.

TA Manifold Pressure is collected on Timer 1

Filaments used: None

SW Version (386): 515 SW Version (EGA): N/A

Sweep and hop mode version(s): None

Super modes and Sweep modes used: None

1) TEGA ON
2) Execute Sequence 30A
3) Select oven (as input parameter) Sol _121____ TA ___0____
4) Activate Sequence 24A
5) Change valve Pulse time to 4mS
6) Set TA_MANIFOLD_PRES to timer_1
7) Set TA OUTLET_PRES to timer 1
8) Loop 100 times (10 second loop)
a) Drive valve open using register write =2049
b) FPGA Dump B to verify valve state.
c) Drive valve closed using register write =4097
d) FPGA Dump B to verify valve state.
9) Start TA Manifold Heater to regulate at 65 C (Heat manifolds to ensure pressure sensors work
10) Start EGA Manifold Heater to regulate at 35 C
11) Loop 100 times (10 second loop)
a) Drive valve open using register write =2049
b) FPGA Dump B to verify valve state.
c) Drive valve closed using register write =4097
12) FPGA Dump B to verify valve state Loop 180 times (5 second loop) aka “A BUNCH”
a) Drive valve open using register write =2049
b) FPGA Dump B to verify valve state.
13) Set timer 5 to 8 s (to control on-off times of inlet or bypass valves for flow regulation)

14) Control bypass valve via duty cycle
15) Pause 520 seconds
16) Stop control of bypass valve
1. Set timer 5 to 16 s (to control on-off times of inlet or bypass valves for flow regulation)

17) Open oven outlet valve
18) Control oven inlet valve via duty cycle
19) Pause1040 seconds
20) Stop control of inlet valve
21) Close outlet valve
22) Delay 10 minutes to determine leak rate
23) Disable heaters
24) Do TEGA TA 4 check with “check_oven=0”
25) Call power off