Power on and perform a full high-temperature ramp on the sample in TA7, which has been previously heated via a mid-temperature ramp. This sample had a plugged frit during the low and mid-T runs, and is assumed to be plugged here. The purpose of this run is to look for the enthalpy associated with decomposition of altered minerals and to measure the gases given off by the decompositions.
Mid-temperature ramp has been completed previously. The carrier valve is closed and the TA manifold has full tank pressure (about 6500 dN). The carrier valve will not be operated due to its erratic behavior.
Software updates were made post Sol 113 to the 386SW and the mode table was changed to V125
Inputs: Multiplier Voltage: __3300 Calibration Set: 3 Ch 4 Calibration Voltage: 2800 Pump Nudge needed: Yes Venting Oven #: Will not vent -(none) Pre-heat duration: 30 secs Calibrate Channels: 1,4 Oven Close Time: NA Select TA #: 7 TA# for Check:
N/A Emission: Low Cal Tank Temperature: -99C
Constraints:Gas(es) used: Mars ATM will be collected through the TA Sample valve. Filaments used: 1 SW Version (386): V515 SW Version (EGA): V321
October
2, 2008 TEGA_High-Temp_TA7_v4_r4.doc
TEGA High-Temp Soil (TA7_plugged) Sweep and hop mode version(s): MODE TABLE V125.
Supermodes and Sweep modes used: Supermodes 1, 2, 3, 4, 6, and 7; Sweeps 37, and 38.
Note: TEGA is checked to see if TEGA is safed. If so, abort self to end activity
10.1. (LOW), Calibration Set 3
10.2. Calls tega_ega_prep, filament-1
10.2.1.1. Sets JPlate 0 to -107V
10.2.1.2. Sets Jplate 1 to -92V
10.2.1.3. Pause 5 seconds
10.2.1.4. Set MV 3300 (input variable)
10.2.1.5. Ion Pump ON
10.2.1.6. Start SM 1, pause 1 sec
10.2.1.7. Emission ON
10.2.1.8. Pause 2 Minutes
10.2.1.9. Stop SM 1
10.2.1.10. Emission ON
10.2.1.11. Pause 2 Minutes
10.2.1.12. Stop SM 1
10.3. Pause 2 minutes (provides time for ion pump to stabilize)
10.4. Change engineering transmit to 10 secs
10.4.1. IF pass thru = 1, emission current will change to HIGH
10.5. Call tega_ega_calibration_scans
10.5.1. calibration_set _____3________
10.5.2. hi_emission_______LOW ____
10.5.3. cal_channels______1,4______
10.5.4. cal_voltage ______3300V ____
10.5.5. cal_4_voltage_____2800V ____
10.5.6. exit_voltage ______3300V ____
| 10.6. | Wait 1 second |
|
|---|---|---|
| 10.7. | Start SM 3 |
|
| 10.8. | Pause 20 minu | tes |
| 10.9. | Stop SM 3 |
|
| 10.10. | Pause 1 minute |
|
| 10.11. | Calls tega_ega10.11.1.1.1. 10.11.1.1.2. 10.11.1.1.3. 10.11.1.1.4. 10.11.1.1.5. 10.11.1.1.6. | _gain Start SM 2 Pause 1 minute Set MV 2600 – 3400V (1 minute per voltage) Exit at 3300_V (input variable) Stop SM 2 Run Sweep 38 |
EGA is warmed up and ready to go (54 mins)
- 3 -
15.1.1. calibration_set _____3________
15.1.2. hi_emission_______LOW ____
15.1.3. cal_channels______1,4______
15.1.4. cal_voltage ______3300V ____
15.1.5. cal_4_voltage_____2800V ____
15.1.6. exit_voltage ______3300V ____
Remember to turn on current monitoring to shut down manifold heater (Do not turn manifold heat back on after turning it off in the ramp.)
Execute Sequence 24 B to increase readings on temperature regulation.
23.1.1. TA Manifold +65C
23.1.2. Plumbing 1 +65C
23.1.3. Plumbing 2 +35C
23.1.4. Transfer Tube +75C
23.1.5. EGA Plumbing +45C
23.1.6. EGA manifold to +35 C
23.1.7. Wedge Heater @ 10% duty cycle
October
2, 2008 TEGA_High-Temp_TA7_v4_r4.doc
TEGA High-Temp Soil (TA7_plugged)
70.1. Turn off the third TEGA switch
70.2. Turn off the second TEGA switch
70.3. Turn off the first TEGA switch
70.4. FPGA Dump – Aker Valve state