October 2, 2008 TEGA_High-Temp_TA7_Day2_v1_r2.doc

TEGA High-Temp Soil (TA7_plugged) DAY 2

Reconciled with block: tega_sporty_ht_ramp_ta7_2 Objectives:

Power on and perform Day-2 of a full high-temperature ramp on the sample in TA7, which has been previously heated via midday-1 High-temperature ramp. This sample had a plugged frit during the low and mid-T runs, and is assumed to be plugged here. The purpose of this run is to provide a baseline for the enthalpy associated with decomposition of altered minerals found on Day-1 and to provide a baseline for the gases given off by the decompositions. We are flowing no carrier gas, so the calling block will be identical to the Day-1 block except that the TA Sample valve is left closed throughout the run.

Configuration:

High-temperature Day-1 ramp has been completed previously. The carrier valve is closed and the TA manifold has no gas in it. It was emptied during the Day-1 procedure. The carrier valve will not be operated due to its erratic behavior. Software updates were made post Sol 113 to the 386SW and the mode table was changed to V125

Inputs:Multiplier Voltage: __3300Calibration Set: 3Ch 4 Calibration Voltage: 2800Pump Nudge needed: Yes Venting Oven #: Will not vent -(none)Pre-heat duration: 30 secsCalibrate Channels: 1,4Oven Close Time: NA Select TA #: 7TA# for Check: N/AEmission: LowCal Tank Temperature: -99C

Constraints:Gas(es) used: Mars ATM will be collected through the TA Sample valve.

October 2, 2008 TEGA_High-Temp_TA7_Day2_v1_r2.doc

TEGA High-Temp Soil (TA7_plugged) DAY 2

Filaments used: 1

SW Version (386): V515 SW Version (EGA): V321

Sweep and hop mode version(s): MODE TABLE V125.

Supermodes and Sweep modes used: Supermodes 1, 2, 3, 4, 6, and 7; Sweeps 37, and 38.

Supermodes used in Ramp: 4, 6, & 7 (and 29, which is a dummy call to not change the supermode

Note: TEGA is checked to see if TEGA is safed. If so, abort self to end activity

    1. Turn ON TEGA – Call TEGA_power_on – This is controlled by the interop science master
    2. 1.1.1. TEGA_MAIN1_PWR "ON"
    3. 1.1.2. TEGA_MAIN2_PWR "ON"
    4. 1.1.3. TEGA_MAIN3_PWR "ON"
    5. 1.1.4. Start the TEGA 386 FSW
  1. Pause 3 Minutes (collect engineering data at a high rate before turning down)
  2. Call Sequence 30A

4. Pre-heat delay 30 secs

    1. Turn on EGA
    2. 5.1.1. TEGA/FSW:EGA:Change EGA State
    3. 5.1.2. Restore the modes from EEPROM 0
    4. 5.1.3. EGA Commands:Restore Mode
  1. OPEN GEC Vent Valve
  2. Activate Sequence 24 A (Turns up readings for regulating temps)
  3. Set Manifold Heater to -30C (execute Sequence 3b)
  4. Set Emission on delay to 5 sec (allows best chance for emission start if large Ion pump current)
  5. Warm up EGA: CALL tega_ega_emission_warmup

10.1. (LOW), Calibration Set 3

10.2. Calls tega_ega_prep, filament-1

10.2.1.1. Sets JPlate 0 to -107V

10.2.1.2. Sets Jplate 1 to -92V

10.2.1.3. Pause 5 seconds

10.2.1.4. Set MV 3300 (input variable)

October

2, 2008 TEGA_High-Temp_TA7_Day2_v1_r2.doc

TEGA High-Temp Soil (TA7_plugged) DAY 2

10.2.1.5. Ion Pump ON

10.2.1.6. Start SM 1, pause 1 sec

10.2.1.7. Emission ON

10.2.1.8. Pause 2 Minutes

10.2.1.9. Stop SM 1

10.2.1.10. Emission ON

10.2.1.11. Pause 2 Minutes

10.2.1.12. Stop SM 1

10.3. Pause 2 minutes (provides time for ion pump to stabilize)

10.4. Change engineering transmit to 10 secs

10.4.1. IF pass thru = 1, emission current will change to HIGH

10.5. Call tega_ega_calibration_scans

10.5.1. calibration_set _____3________

10.5.2. hi_emission_______LOW ____

10.5.3. cal_channels______1,4______

10.5.4. cal_voltage ______3300V ____

10.5.5. cal_4_voltage_____2800V ____

10.5.6. exit_voltage ______3300V ____

10.6. Wait 1 second
10.7. Start SM 3
10.8. Pause 20 minu tes
10.9. Stop SM 3
10.10. Pause 1 minute
10.11. Calls tega_ega10.11.1.1.1. 10.11.1.1.2. 10.11.1.1.3. 10.11.1.1.4. 10.11.1.1.5. 10.11.1.1.6. _gain Start SM 2 Pause 1 minute Set MV 2600 – 3400V (1 minute per voltage) Exit at 3300_V (input variable) Stop SM 2 Run Sweep 38

EGA is warmed up and ready to go (54 mins)

11. Reinforce Emission on

(IF pump nudge is YES Steps 12 to 14 are followed, if NO skip to 15)

  1. Change to HIGH Emission
  2. Pause 5 minutes

- 3 - October

2, 2008 TEGA_High-Temp_TA7_Day2_v1_r2.doc

TEGA High-Temp Soil (TA7_plugged) DAY 2

  1. Change to LOW Emission
    1. Calibrate EGA: CALL tega_ega_calibration_scans
    2. 15.1.1. calibration_set _____3________
    3. 15.1.2. hi_emission_______LOW ____
    4. 15.1.3. cal_channels______1,4______
    5. 15.1.4. cal_voltage ______3300V ____
    6. 15.1.5. cal_4_voltage_____2800V ____
    7. 15.1.6. exit_voltage ______3300V ____
  2. Set EGA manifold heater to +35 C
  3. Set TA manifold heater to +65 C

18. Set

following readings to be taken every 1 second.

18.1. Manifold Pressure

18.2. Outlet Pressure

18.3. Bus A Current

18.4. Bus B Current

    19.
    Start Supermode 7: Soil volatiles
  1. 20. Select Oven TA_7__ (TWEAKABLE)
  2. 21.
    Set Ramp so it has no control over any of the valves(6 total)

Remember to turn on current monitoring to shut down manifold heater (Do not turn manifold heat back on after turning it off in the ramp.)

      22.
      Start Ramp (TWEAKABLE)
    1. 22.1. Ramp Segments File:_______________________________________
  1. 23. Run Heater Init (tega_init_heater_control)

Execute Sequence 24 B to increase readings on temperature regulation.

23.1.1. TA Manifold +65C

23.1.2. Plumbing 1 +65C

23.1.3. Plumbing 2 +35C

23.1.4. Transfer Tube +75C

23.1.5. EGA Plumbing +45C

23.1.6. EGA manifold to +35 C

23.1.7. Wedge Heater @ 10% duty cycle

24. Pause for Delay_2 (467 sec;) – TWEAKABLE

October

2, 2008 TEGA_High-Temp_TA7_Day2_v1_r2.doc

TEGA High-Temp Soil (TA7_plugged) DAY 2

Measure Atmosphere

.

    1. Pause for Delay_3 (879 sec;) – TWEAKABLE TA Manifold = 4000 dN = 66 mbar Flush atmosphere from oven and plumbing (assumes oven is not plugged at low T)
      1. 26.
        Open oven inlet valve
        27.
        Pause 2 minutes (to measure leak rate of oven) TA Manifold = 18 mbar
        28.
        Open oven outlet valve (to flush oven)
      2. 29. Pause 10 seconds (to get an idea on flow rate)
  1. Open bypass valve (to flush atmosphere out of EGA manifold)
  2. 31. Pause 2 seconds
  3. Close bypass valve
  4. Close outlet valve (inlet remains open so TA manifold can record any pressure rise)
    1. 34. Pause for Delay_4 ( 2031 sec for 650 deg C ) - TWEAKABLE Take gas sample at 650 C temperature.
      1. Open oven outlet valve (to flush oven)
      2. Pause 10 seconds (to give gas a chance to escape in case it is not plugged)
  1. Open bypass valve (to let evolved gases get to EGA outlet manifold)
  2. Pause 2 seconds
  3. Close bypass valve
  4. Close outlet valve
  5. 41. Pause 2 minutes
  6. 42.
    Pause 1 min (to measure transition with high time resolution SM 6)
  7. 43. Stop SM

October

2, 2008 TEGA_High-Temp_TA7_Day2_v1_r2.doc

TEGA High-Temp Soil (TA7_plugged) DAY 2

44. Start SM 7

49. Pause 1 min (to watch gas drain out of CML)

Shut down

  1. Run tega_disable_heater_control (Turn all remaining heaters off)
  2. Close bypass valve

October

2, 2008 TEGA_High-Temp_TA7_Day2_v1_r2.doc

TEGA High-Temp Soil (TA7_plugged) DAY 2

62. CLOSE GEC Vent Valve Formatted: Bullets and Numbering

    1. Turn off EGA: CALL tega_ega_off Formatted: Bullets and Numbering
    2. 63.1. EGA Shutdown:
      63.1.1. Multiplier Voltage OFF
      63.1.2. Wait 5 secs
      63.1.3. Emission Off
      63.1.4. Wait 5 secs
      63.1.5. Ion Pump Control OFF
      63.1.6. Activate Sequence 7A to turn the EGA OFF
  1. Turn off TEGA: CALL tega_power_off

64.1. Turn off the third TEGA switch

64.2. Turn off the second TEGA switch

64.3. Turn off the first TEGA switch

64.4. FPGA Dump – Aker Valve state

Open TA sample valve (to reduce pressure to ambient in CML volume) Close TA sample valve