Reconciled with block: Uplinkin Sol 141
Objectives:
Measure the composition of the atmosphere with special emphasis on CO2 isotopes. We will measure the atmosphere via the TA Sample inlet, Cal Gas via the TA Sample inlet, and then finally atmosphere via both inlets.
Configuration:
The carrier valve is not too reliable, but it will not be used here. Just in case, we will open and close valves with multiple commands.
Software updates were made post Sol 113 to the 386SW and the mode table was changed to V125
Note: Step 55 can be reduced to 5 mins, Step 57 can be reduced to 5 minutes if total duration is a problem for power.
Inputs:
Multiplier Voltage: __3300
Calibration Set: 3
Ch 4 Calibration Voltage: 2800
Pump Nudge needed: Yes
Venting Oven #: Will not vent - (none)
Pre-heat duration: 30 secs
Calibrate Channels: 1,4
Oven Close Time: NA
Select TA #: NA
TA# for Check: N/A
Emission: Low
Cal Tank Temperature: -99 C
Constraints:
Gas(es) used: Mars ATM will be collected through the TA Sample valve.
Filaments used: 1
SW Version (386): V515 SW Version (EGA): V321
Sweep and hop mode version(s): MODE TABLE V125.
Supermodes and Sweep modes used: Supermodes 1, 2, 3, 4, and 9; Sweeps 37, and 38.
Supermodes used in Ramp: No ramp to be run
Note: TEGA is checked to see if TEGA is safed. If so, abort self to end activity
-
1. Turn ON TEGA – Call TEGA_power_on – This is controlled by the interop science master
-
1.1.1. TEGA_MAIN1_PWR "ON"
-
1.1.2. TEGA_MAIN2_PWR "ON"
-
1.1.3. TEGA_MAIN3_PWR "ON"
-
1.1.4. Start the TEGA 386 FSW
-
2. Pause 2 Minutes (collect engineering data at a high rate before turning down)
-
3. Call Sequence 30A
-
4. Pre-heat delay 30 secs
-
5. Turn on EGA
-
5.1.1. TEGA/FSW:EGA:Change EGA State
-
5.1.2. Restore the modes from EEPROM 0
-
5.1.3. EGA Commands:Restore Mode
-
-
6. OPEN GEC Vent Valve
-
7. Activate Sequence 24 A (Turns up readings for regulating temps)
-
8. Set Manifold Heater to -20C (execute Sequence 3b)
-
9. Set EGA Manifold heater to 35C
-
10. Set Emission on delay to 5 sec (allows best chance for emission start if large Ion pump current)
-
11. Warm up EGA: CALL tega_ega_emission_warmup
-
11.1. (LOW), Calibration Set 3
-
-
11.2. Calls tega_ega_prep, filament-1
-
11.2.1.1. Sets JPlate 0 to -107V
-
11.2.1.2. Sets Jplate 1 to -92V
-
11.2.1.3. Pause 5 seconds
-
11.2.1.4. Set MV 3300 (input variable)
-
11.2.1.5. Ion Pump ON
-
11.2.1.6. Start SM 1, pause 1 sec
-
11.2.1.7. Emission ON
-
11.2.1.8. Pause 2 Minutes
-
11.2.1.9. Stop SM 1
-
11.2.1.10. Emission ON
-
11.2.1.11. Pause 2 Minutes
-
11.2.1.12. Stop SM 1
-
11.3. Pause 2 minutes (provides time for ion pump to stabilize)
-
11.4. Change engineering transmit to 10 secs
-
11.4.1. IF pass thru = 1, emission current will change to HIGH
-
11.5. Call tega_ega_calibration_scans
-
11.5.1. calibration_set _____3________
-
11.5.2. hi_emission_______LOW ____
-
11.5.3. cal_channels______1,4______
-
11.5.4. cal_voltage ______3300V ____
-
11.5.5. cal_4_voltage_____2800V ____
-
11.5.6. exit_voltage ______3300V ____
-
11.6. Wait 1 second
-
11.7. Start SM 3
-
11.8. Pause 20 minutes
-
11.9. Stop SM 3
-
11.10. Pause 1 minute
-
11.11. Calls tega_ega_gain
-
11.11.1.1.1. Start SM 2
-
11.11.1.1.2. Pause 1 minute
-
11.11.1.1.3. Set MV 2600 – 3400V (1 minute per voltage)
-
11.11.1.1.4. Exit at 3300_V (input variable)
-
11.11.1.1.5. Stop SM 2
-
11.11.1.1.6. Run Sweep 38
EGA is warmed up and ready to go (54 mins)
-
12. Reinforce Emission on
(IF pump nudge is YES Steps 12 to 14 are followed, if NO skip to 15)
-
13. Change to HIGH Emission
-
14. Pause 3 minutes
-
15. Change to LOW Emission
-
16. Calibrate EGA: CALL tega_ega_calibration_scans
-
16.1.1. calibration_set _____3________
-
16.1.2. hi_emission_______LOW ____
-
16.1.3. cal_channels______1,4______
-
16.1.4. cal_voltage ______3300V ____
-
16.1.5. cal_4_voltage_____2800V ____
-
16.1.6. exit_voltage ______3300V ____
-
17. Start Supermode 9
Measure residuals for 7 minutes
-
18. Pause for 7 minutes
Measure Atmosphere (via TA Sample inlet)
-
19. Begin SM 4
-
20. Pause 30 seconds
-
21. Open TA Sample valve to measure atmosphere.
-
22. Pause 2 minutes
-
23. Close TA Sample valve to capture atmosphere
-
24. Open bypass valve
-
25. Pause 30 seconds
-
26. Start SM 9 (To collect the high-quality atmospheric data)
Begin measurement of Cal Gas
-
27. Regulate Cal Gas to 35 mbar.
-
28. Pause 12 minutes (to allow cal gas to flush out manifold and to finish analysis of atmosphere)
-
29. Start SM 4
-
30. Pause 30 sec
-
31. Open TA Sample valve (We have been analyzing atmosphere up to this point)
-
32. Pause 2 min
-
33. Close TA Sample valve (to capture Cal Gas)
-
34. Pause 30 seconds
-
35. Start SM 9
-
36. Pause 10 minutes
-
37. Regulate Cal Gas to 25 mbar
-
38. Open TA Sample valve
-
39. Pause 2 minutes
-
40. Close TA Sample valve
-
41. Close bypass valve
-
42. Stop Cal Gas Regulation
-
43. Poke close with dump 5 sec delay, 10 cycles
-
44. Pause 10 minutes (to let cal gas begin to diffuse out of EGA manifold and to complete analysis of Cal gas)
Measure atmosphere via Atm inlet, and empty cal gas from EGA manifold.
-
45. Start SM 4
-
46. Pause 30 seconds
-
47. Open TA Sample valve
-
48. Pause 5 minutes (to be sure most Cal gas is gone from EGA manifold)
-
49. Close TA Sample valve
-
50. Pause 30 seconds
-
51. Start SM 9
-
52. Pause 5 minutes (to get reading on composition of gas in TA Sample CML)
-
53. Open Atmosphere inlet valve
-
54. Pause 10 minutes (to get data without decay of pressure in CML volume)
-
55. Poke close ATM Valve with dump 5 sec delay, 10 cycles
-
56. Close Atmosphere inlet valve
-
57. Pause 20 minutes (to get data with decay of gas in CML volume, to see if isotopic ratio is a function of count rate; time is TBD depending on what we can get.)
Shut down
-
58. Run tega_disable_heater_control (Turn all remaining heaters off)
-
59. Stop SM
-
60. CLOSE GEC Vent Valve
-
61. Turn off EGA: CALL tega_ega_off
-
61.1. EGA Shutdown:
-
61.1.1. Multiplier Voltage OFF
-
61.1.2. Wait 5 secs
-
61.1.3. Emission Off
-
61.1.4. Wait 5 secs
-
61.1.5. Ion Pump Control OFF
-
61.1.6. Activate Sequence 7A to turn the EGA OFF
-
62. Turn off TEGA: CALL tega_power_off
-
62.1. Turn off the third TEGA switch
-
62.2. Turn off the second TEGA switch
-
62.3. Turn off the first TEGA switch
-
62.4. FPGA Dump – Aker Valve state